Chapter 3
Maintaining a Large Volume Splitless Injector
Replace a Broken Liner
68
Maintenance and Troubleshooting Manual
5. When the inlet reaches the room temperature,
press
LEFT CARRIER
or
RIGHT CARRIER
depending on which injector is operating.
RIGHT CARRIER (S/SL)
Pressure
30.0
30.0
Col. flow
3.00
<
Lin. Veloc.
(60.9)
6. Scroll to
Pressure
and press
OFF
.
7. Unscrew the injector cap.
8. Unscrew the ring nut and remove the septum
holder with the septum from the inlet port.
9. Use the liner cap wrench provided with the GC
to remove the injector liner cap.