Chapter 15
Maintaining a Photoionization Detector
PID Troubleshooting
336
Maintenance and Troubleshooting Manual
Symptom
Diagnosis
Remedy
Tailing of sample peaks
Adsorption effects caused by
degradation products inside
the cell
Remove deposits by baking the cell
for two hours at 400°C.
If the symptom does not disappears,
contact your customer support
organization.
Refer to Appendix B,
, for contact
information.
Contamination of the cell
Remove contaminants by baking the
cell for two hours at 400°C.
If the symptom does not disappears,
contact your customer support
organization.
Refer to Appendix B,
, for contact
information.
Inadequate gas flow rate
settings
Set gas flow rates to proper values.
Leaks on the gas lines
Perform a leak test.
Refer to the
Low sensitivity
Lamp window fogged by UV-
absorbing deposits
Clean the lamp window.
Refer to the
Operating Sequence on page 323.
Flow rate of the make-up gas
is set too high
Reduce the flow rate of the make-up
gas down to10 ml/min or less.
UV lamp is not adequate to
the compounds to be analyzed
Replace with an appropriate one.
Table 15-3.
PID troubleshooting table (Continued)