Chapter 11
Maintaining a FID
Replace the Collecting Electrode
208
Maintenance and Troubleshooting Manual
4. Disconnect the signal and ignition cables from
the detector.
5. Loosen the fixing nut on the front of the
detector cell and remove it.
6. Unscrew and remove the cap of the detector
cell.
7. Remove the holder assembly of the signal cable
by unscrewing the two fixing screws.
8. Using forceps, unscrew and remove the gold
contact pin.
9. Using tweezers, extract the collecting electrode
through the top of the detector cell.