Chapter 9
Maintaining a PTV Injector
Replace a Broken Liner
162
Maintenance and Troubleshooting Manual
17. Tighten the liner cap using the screwdriver
provided with the GC.
18. Insert the analytical column with its ferrule into
the bottom of the injector in its previous
position. Refer to the TRACE
™
GC Operating
Manual for instructions.
NOTE
In split and splitless injection
modes, the distance
d
between
the top of the capillary column and
the ferrule must be 3 cm.
If you are using a
simile on-
column
liner, in order to operate in
on-column injection mode, insert
the column as far as it goes.
19. Tighten the M4 retaining nut to hold the column
in place.