Chapter 17
Maintaining Detectors in Stack Configuration
Clean or Replace the Collecting Electrode (NPD on ECD)
356
Maintenance and Troubleshooting Manual
OPERATING SEQUENCE
Clean or Replace the Collecting Electrode (NPD on ECD)
Materials needed
•
Screwdriver
•
2.5 mm Allen wrench
•
3 mm Allen wrench
•
Tweezers
for replacement:
•
Collecting electrode (P/N 277 003 50)
for cleaning:
•
Ultrasonic cleaning bath
•
GC-grade hexane
•
GC-grade toluene
•
Fine emery paper (optional)
WARNING! Any maintenance or repair operations involving the radioactive source of the ECD
must be performed ONLY by qualified personnel duly authorized to handle
radioactive material. Please contact your local Thermo Fisher Scientific Technical
Service office.
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
for more information.
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detectors must be cooled to room temperature.
PRECAUTIONS
Содержание TRACE GC Ultra
Страница 4: ......
Страница 10: ...x Maintenance and Troubleshooting Manual Contents ...
Страница 20: ......
Страница 128: ......
Страница 178: ...Chapter 9 Maintaining a PTV Injector PTV Troubleshooting 178 Maintenance and Troubleshooting Manual ...
Страница 218: ......
Страница 268: ......
Страница 314: ......