Chapter 19
Peaks
Analytical Troubleshooting
Maintenance and Troubleshooting Manual
397
Behavior
Characteristics
Cause
Remedy
Negative
peaks
Some
Symptom can be normal
No peaks
after solvent peak
Carrier gas flow too high
Reduce the carrier gas flow rate.
Combustion gas flow
incorrect
Check the combustion gas flow.
Detector contaminated
Bake out or clean the detector.
FID flame extinguished by
solvent peak
Check the detector temperature.
Too much sample injected
Inject less sample.
Incorrect column position
in S/SL injector (too high)
Check the column position.
No peaks
at all
Clogged syringe needle
Replace or repair the syringe.
Column broken or
disconnected
Check the column and
connections.
Defective electrometer or
amplifier
Replace the electrometer or
amplifier.
Defective recording device
Replace the recording device.
FID flame is out
Light the flame.
Poor or missing electrical
connection
Check the cable connections.
Incorrect column position
in S/SL injector (too high)
Check the column position.
Tailing
Sample peaks
Column degradation
causing activity
Inject a test mixture and evaluate
the column.
Column/oven temperature
too low
Increase the column/oven
temperature. Do not exceed the
recommended maximum
temperature for the stationary
phase.
Dirty liner
Clean or replace the liner.