Chapter 14
Clean the Jet
Maintaining a Flame Photometric Detector
Maintenance and Troubleshooting Manual
275
6. Using the tool provided with the GC, loosen the
fixing nut on the base of the detector and
remove it.
CAUTION
Pay attention not to lose the
aluminium ring inserted
between the detector head and
the base body.
7. Using the 5 mm wrench provided with the GC,
loosen the jet and remove it from the detector
base body.
8. Place the jet in the ultrasonic bath filled with
liquid detergent and clean it for about five
minutes.
9. Handling the jet with forceps or tweezers, rinse
the jet with distilled water, then with methanol.
10. Place the jet on a paper towel and let it air dry.
PRECAUTIONS