Chapter 17
Replace the Silver Seal (FID on ECD)
Maintaining Detectors in Stack Configuration
Maintenance and Troubleshooting Manual
369
OPERATING SEQUENCE
Replace the Silver Seal (FID on ECD)
Materials needed
•
Silver seal (P/N 290 150 58)
•
ECD fixing tool (P/N 205 021 50)
•
Tweezers
WARNING! Any maintenance or repair operations involving the radioactive source of the ECD
must be performed ONLY by qualified personnel duly authorized to handle
radioactive material. Please contact your local Thermo Fisher Scientific Technical
Service office.
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detectors must be cooled to room temperature.
PRECAUTIONS
1.
.
AUXILIARY
Signal
<
Detector
Temperature Zones
Pressure Channel
Press
AUX
.
2. Scroll to
Detector
and press
ENTER
.
Содержание TRACE GC Ultra