Chapter 9
Maintaining a PTV Injector
Adjust a Quartz Wool Packing
168
Maintenance and Troubleshooting Manual
12. Insert a thin, non metallic bar into the ends of
the liner and push the quartz wool so to have a
distance of about 28 mm between the packing
and the ends of the liner.
CAUTION
Do not compress the wool, to
avoid carrier gas cannot flow
through the liner.
13. Place a new graphite seal over the liner, making
sure to leave a distance of about 10 mm
between the seal and the liner end.
14. Using tweezers, insert the liner into the injector
and push it gently towards the bottom fitting.
CAUTION
Be careful not to damage the
graphite seal or allow graphite
to entering the liner. Should this
occur, clean the liner with an
inert gas.
PRECAUTIONS