Chapter 9
Clean or Replace the Liner
Maintaining a PTV Injector
Maintenance and Troubleshooting Manual
153
9. Use the
screwdriver provided with the GC to
remove the injector liner cap.
10. Use tweezers to remove the liner with the
graphite seal.
CAUTION
If you are using a glass liner, be
careful not to break it inside the
injector when removing it. Glass
splinters might fall into the
lower part of the vaporization
chamber. If the liner breaks,
refer to the
Operating Sequence on
11. If you are going to replace the liner with a new
one, go to step 13. If you are going to clean the
liner, put the dirty liner into an ultrasonic bath
filled with a methanol/acetone mixture (1:1)
and sonicate it for about half an hour.