Chapter 9
Maintaining a PTV Injector
Replace a Broken Liner
160
Maintenance and Troubleshooting Manual
11. Unscrew the nut that retains the analytical
column.
12. Remove the analytical column with its ferrule.
13. Inserting the steel thread provided specifically
for this purpose, remove the possible glass
fragments from the vaporization chamber
pushing them out through the upper opening of
the injector.
CAUTION
The spacer placed below the
liner will also be ejected from
the injector: pay attention not to
lose it.