Chapter 11
Replace the Ignition Assembly
Maintaining a FID
Maintenance and Troubleshooting Manual
213
7. Remount the new connector holder assembly
and fix it to the detector cell using the two
screws. Make sure the connecting fork is placed
downwards to achieve the correct contact with
the jet.
8. Place the detector cell on the detector base body
and tighten the retaining screw.
9. Reconnect the signal and ignition cables to the
detector.
10. Reconnect the signal and ignition cables to the
detector.
11. Reset the detector to the required operating
conditions. Refer to the relevant chapter of the
TRACE
TM
GC Ultra
Operating manual
for
more information.