Chapter 14
Replace the Jet
Maintaining a Flame Photometric Detector
Maintenance and Troubleshooting Manual
295
8. Place a new jet into the housing and tighten it.
9. Place the detector on its base body, paying
attention that the aluminium ring has been
inserted in the correct position, then tighten the
fixing nut.