Chapter 3
Replace a Broken Liner
Maintaining a Large Volume Splitless Injector
Maintenance and Troubleshooting Manual
69
10. Use tweezers to remove the upper part of the
broken liner (with the graphite seal) from the
injector.
11. Unscrew the nut that retains the analytical
column.
12. Remove the analytical column with its ferrule.