Chapter 15
Maintaining a Photoionization Detector
Clean the Lamp Window
324
Maintenance and Troubleshooting Manual
1. Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the PID detector,
to access the PID control table. In the following
example, a PID installed on the right channel is
considered.
.
RIGHT DET (PID)
Lamp
Off <
Base temp
250
250
High current
N
Signal pA
(xx)
Mkup(N2)
7
Off
Sheath gas
40
Off
2. Scroll subsequently to
Lamp
,
Mkup
and
Sheath
Gas
press
OFF
.
3. Loosen the two knurled screws that fix the lamp
holder and remove it from the lamp housing.