Chapter 14
Clean the Mirror Metal Plug
Maintaining a Flame Photometric Detector
Maintenance and Troubleshooting Manual
283
1. Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the FPD detector,
to access the FPD control table. In the
following example, a FPD installed on the right
channel is considered.
.
RIGHT DET (FPD)
Flame
Off <
Base temp
300
Off
FPD temp
150
Off
Signal pA
(xx)
High voltage mode?
N
H2
90
Off
Air
115
Off
Mkup(N2)
0
0
2. Scroll to
Flame
and press
OFF
.
3. Scroll subsequently to
Base temp,FPD
temp, H2
and
Air
and press
OFF
.
4. Switch off the GC power supply.
WARNING! Before disconnecting the cables
or disassembling the detector,
switching off the GC power
supply is indispensable to avoid
electrical shocks and damages
to the instrument.
5. Disconnect the signal, excitation voltage and
ignition/heating cables from the detector.