Chapter 3
Maintaining a Large Volume Splitless Injector
Clean or Replace the Liner
62
Maintenance and Troubleshooting Manual
3. Press
LEFT INLET
or
RIGHT INLET
depending on
which injector is operating. In the following
example, a S/SL injector installed on the right
channel is considered.
RIGHT INLET (S/SL)
Temp
250
250 <
Pressure
100
100
Mode
split
4. Scroll to
Temp
and press
OFF
.
5. When the inlet reaches the room temperature,
press
LEFT CARRIER
or
RIGHT CARRIER
depending on which injector is operating.
RIGHT CARRIER (S/SL)
Pressure
30.0
30.0
Col. flow
3.00
<
Lin. Veloc.
(60.9)
6. Scroll to
Pressure
and press
OFF
.
7. Unscrew the injector cap.
8. Unscrew the ring nut and remove the septum
holder with the septum from the inlet port.