Chapter 17
Clean or Replace the Collecting Electrode (NPD on ECD)
Maintaining Detectors in Stack Configuration
Maintenance and Troubleshooting Manual
357
1.
.
AUXILIARY
Signal
<
Detector
Temperature Zones
Pressure Channel
Press
AUX
.
2. Scroll to
Detector
and press
ENTER
.
3.
.
AUX DETECTOR (NPD)
Source cur,A
2.740 <
Base temp
300
Off
Signal pA
(10.2)
Polarizer V
3.5
H2 delay time
Off
H2
2
2
Air
60
60
In the
AUX DETECTOR (NPD)
table, scroll to
Base temp
and press
OFF
.
4. Scroll subsequently to
Source curr,H2
,
Air
and press
OFF
.
5. Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the ECD, to access
the ECD control table.
.
XXX DETECTOR (ECD)
Base temp
250
Off <
ECD temp
250
Off
Ref current nA
1.0
Freq kHz
(2.15)
Pulse amp V
25
Pulse width us
1.0
Mkup (N2)
40
Off
6. Scroll subsequently to
Base temp
and
ECD
temp
and press
OFF
.
7. Scroll to
Mkup(N2)
and press
OFF
.