Chapter 12
Maintaining an Electron Capture Detector
Replace the Silver Seal
234
Maintenance and Troubleshooting Manual
8. Insert a new silver seal on the lower part of the
ECD detector.
9. Remount the detector on its base body,
tightening the fixing nut with the proper tool.
10. Reconnect the signal and ignition cables to the
detector.
11. Reset the detector to the required operating
conditions.