Chapter 14
Maintaining a Flame Photometric Detector
Replace the Heat Shields
306
Maintenance and Troubleshooting Manual
16.
5 mm
Insert the spacer into the flange, letting it stand
out for about 5 mm.
17. Insert a new graphite seal on the spacer,
pushing it slowly until it touches the flange.
18. Insert the flange into the detector body and fix it
tightening the three Allen screws.
19. Using a clean paper towel, clean the mirror
surface of the plug. If necessary, use a solvent
as methylene chloride or methanol to remove
deposits and a metal polishing paste to restore it
to the previous reflectivity.