Chapter 16
TCD Troubleshooting
Maintaining a Thermal Conductivity Detector
Maintenance and Troubleshooting Manual
343
Symptom
Diagnosis
Remedy
Low sensitivity
Leaks on the gas lines
Check the connections of the carrier,
reference and make-up gas are tight.
Leak due to septum wearing
Replace the septum.
Operating conditions of the
detector not properly set
Optimize working parameters of the
detector according to the actual
operating mode (Constant Voltage,
Constant Temperature).
Refer to the TRACE GC Ultra
Operating Manual
for instructions.
Thermal conductivity of the
carrier gas is too close to the
one of the compound to be
analyzed
Use a different carrier gas.
Refer to the TRACE GC Ultra
Operating Manual
for instructions.
Contaminated filaments
Remove contamination by baking the
filaments for one hour at a
temperature higher than the boiling
point of the most high-boiling
compound.
Perform baking twice if necessary.
If the symptom does not disappears,
contact your customer support
organization.
Refer to Appendix B,
, for contact
information.
Table 16-2.
TCD troubleshooting table (Continued)