Chapter 12
Replace the Collecting Electrode
Maintaining an Electron Capture Detector
Maintenance and Troubleshooting Manual
231
7. Using tweezers, replace the collecting electrode
with a new one and tighten it.
8. Remount the cover with the chimney on the
detector.
9. Reconnect the signal and ignition cables to the
detector.
10. Reset the detector to the required operating
conditions.