Chapter 15
Clean the Lamp Window
Maintaining a Photoionization Detector
Maintenance and Troubleshooting Manual
323
OPERATING SEQUENCE
Clean the Lamp Window
Materials needed
•
Lamp Polishing kit (P/N 190 048 00)
•
GC-grade methanol or acetone
•
Paper towels
•
Cotton gloves
WARNING! UV radiation can cause injuries to your eyes. Therefore, before beginning the
sequence, make sure the UV lamp has been switched off.
PRECAUTIONS
WARNING! The cleaning of the lamp window can be carried out without removing the detector
from its base body. Nevertheless, keep in mind that parts of the detector are hot and
could cause injuries to your hands.
PRECAUTIONS
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
for more information.