background image

Revision G     Issued November 2013

 Getting Started Guide

DHR Series and AR Series

Small Angle Light-Scattering (SALS)

Accessory

Содержание DHR Series

Страница 1: ...Revision G Issued November 2013 Getting Started Guide DHR Series and AR Series Small Angle Light Scattering SALS Accessory...

Страница 2: ...overing subject matter in this document Except as expressly provided in written license agreement from TA Instruments the furnishing of this document does not give you any license to these patents tra...

Страница 3: ...ruments Manual Supplement Please click the TA Manual Supplement link to access the following important information supplemental to this Getting Started Guide TA Instruments Trademarks TA Instruments P...

Страница 4: ...2 No 61010 2 010 Particular requirements for laboratory equipment for the heating of materials For European Economic Area In accordance with Council Directive 2006 95 EC of 12 December 2006 on the har...

Страница 5: ...Generators For the European Economic Area In accordance with Council Directive 2004 108 EC of 15 December 2004 on the approximation of the laws of the Member States relating to electromagnetic compati...

Страница 6: ...ce and repair of internal parts must be performed only by TA Instruments or other qualified service personnel AVERTISSEMENT Pr sence de tensions lev es dans cet instrument La maintenance et la r parat...

Страница 7: ...f the instrument This manual is intended to be used in conjunction with the DHR Getting Started Guide and the AR G2 AR 2000ex AR 1500ex Rheometer Getting Started Guide For full details on the operatio...

Страница 8: ...Plane of Focus 13 Polarization 14 Intensity Adjustment 14 Intensity Corrections 14 Specifications 15 Chapter 2 Installing the SALS Accessory 16 Installing the SALS Accessory 16 Mounting the Upper and...

Страница 9: ...Accessory Getting Started Guide Page 9 Chapter 4 Using the SALS Accessory 32 Overview 32 Loading the Sample 32 Setting the Focal Position 33 Locking Exposure 33 Selecting the Image Format 33 Exporting...

Страница 10: ...allows the laser to be trained on the sample The resultant scattered light is transmitted through the upper transparent quartz geometry and is observed by a lens and camera assembly The camera images...

Страница 11: ...d for Bragg s law is half the angle defined for SALS by convention the symbol is used rather than for Bragg s law but in this context that would be confusing So L 2 sin 2 sin 2 So L 2 q Thus the great...

Страница 12: ...1024 pixel size 5 2 m square Figure 4 SALS Accessory components Figure 4 above provides a schematic of the SALS accessory components A schematic of the optical sys tem is shown in Figure 5 The pinhol...

Страница 13: ...ich the deflection through 90 is ignored this does not affect principle of the optical system Figure 6 Simplified view of the SALS optics The maximum scattering angle will also depend on the position...

Страница 14: ...for level 3 experiments see Chapter 3 The scattered light falls on a flat sensor consisting of an array of pixels The absolute intensity of the light Iabs is obtained from the intensity reported by t...

Страница 15: ...ng Angle 6 to 26 8 Laser 635nm 0 95mW class 2 Peltier Temperature Controlled 5 to 95 C WARNING The upper temperature limit is set by the optical adhesive that retains the sapphire window Do no exceed...

Страница 16: ...l 2 Attach the upper fixture to the mounting ring on the underside of the instrument head using the three captive screws provided NOTE The camera should project to the right of the instrument when vie...

Страница 17: ...g in the camera USB cable 1 Download the LuCam Software V6 0 3 or higher from the link below http www lumenera com support downloads industrial downloads php 2 Run the downloadable executable LuCamSof...

Страница 18: ...diminish in intensity as the scattering angle increases Mie scattering Polystyrene micro particles of nominal diameter 3 m are suitable for this purpose and these can be obtained from Sigma Aldrich Fl...

Страница 19: ...lower fixture See Figure 10 below Figure 10 Removing the filter 9 Place two or three drops of the alignment solution on the window in the Peltier Plate 10 Turn on the bearing lock and close the geomet...

Страница 20: ...he below right image Figure 11 Swivel mount NOTE It may be necessary to adjust the camera settings such as exposure brightness and contrast by accessing the Camera properties on the camera setup scree...

Страница 21: ...xperiment will determine the choice of format Level 1 Experiments Images can be collected together with the rheological data and viewed in the Data Analysis package Images can be collected in JPEG for...

Страница 22: ...ut variables such as the lens and sensor positions are not known sufficiently precisely to make this a reliable way of determining that relationship Mie Scattering A more accurate method of obtaining...

Страница 23: ...nately the program was not designed to be used with SALS systems so the option of plotting scattering intensity I against q value is not available This means that q values have to be calculated using...

Страница 24: ...l not be smooth If it is set too low the calculation will be excessively time consuming A value of 0 1 gives a reasonable compromise Drop Size Diameter m Enter the value for Calibrated Particle Diamet...

Страница 25: ...t is best to record several images for later averaging It is critical that the focussing and cam era exposure are not adjusted between recording the Mie and water images To prepare images for manipula...

Страница 26: ...ering image in ImageJ Once you have both averaged images the background is subtracted using the Image Calculator window shown below selected from the Process menu Figure 17 Image Calculator window in...

Страница 27: ...ke full advantage of the camera chip this should be set to 512 The integration angle is on both sides of the starting angle so the full angle is twice that entered To integrate over the whole plot use...

Страница 28: ...ith increasing radius but further corrections need to be made for the sensor properties and solid angle This can be done using Microsoft Excel for example Figure 19 Intensity against pixel radius plot...

Страница 29: ...m calibration factors are determined using a worked example in Excel This section shows a worked example using Microsoft Excel Laven data and Intensity vs Radius pixels date from ImageJ Open the txt f...

Страница 30: ...Also add the optics factor at G5 This value together with the y shift is approximate and will be modified later Figure 21 Adjusted data J K L M N O P 1 y shift Radius pixels Intensity Angle Icorr Cor...

Страница 31: ...maximum Adjust the Optics Factor so that the first peaks of each curve overlap The quality of the overlap of the second peak will depend on the dilution factor of the spheres solution See the figure...

Страница 32: ...sam ple to the edge of the upper plate But it is important when performing a SALS experiment that there are no entrained bubbles in the sample as these will tend to disrupt the scattering pattern Unfo...

Страница 33: ...e adjustable front lens posi tions taken from the scale Locking Exposure Exposure as well as other camera settings are made by accessing the camera Settings on the Image panel in TRIOS Help Selecting...

Страница 34: ...ng 25 instrument symbols 6 L license agreement 3 loading the sample 32 locking exposure 33 N notes 4 P patents 3 R Regulatory Compliance 4 S safety 6 instrument symbols 6 Safety Standards 4 SALS align...

Страница 35: ...arted Guide Page 35 preparing and loading calibration sample 25 product description 12 setting the focal position 33 specs 15 using 32 scattering theory 10 setting the focal position 33 T TA Instrumen...

Отзывы: