Chapter 14
Maintaining a Flame Photometric Detector
Clean the Interferential Filter
278
Maintenance and Troubleshooting Manual
OPERATING SEQUENCE
Clean the Interferential Filter
Materials needed
•
Detector fixing tool (P/N 205 021 50)
•
Paper towels
•
GC-grade methanol
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
for more information.
1. Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the FPD detector,
to access the FPD control table. In the
following example, a FPD installed on the right
channel is considered.
.
RIGHT DET (FPD)
Flame
Off <
Base temp
300
Off
FPD temp
150
Off
Signal pA
(xx)
High voltage mode?
N
H2
90
Off
Air
115
Off
Mkup(N2)
0
0
2. Scroll to
Flame
and press
OFF
.
3. Scroll subsequently to
Base temp,FPD
temp, H2
and
Air
and press
OFF
.