background image

INSTRUCTION MANUAL

Foodcare

Wireless pH Tester 

for Beer

with built-in specialized electrode

HI9810312

Содержание HALO 2

Страница 1: ...INSTRUCTION MANUAL Foodcare Wireless pH Tester for Beer with built in specialized electrode HI9810312...

Страница 2: ...a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument a...

Страница 3: ...ffer Set 6 pH Resolution 6 Bluetooth Wireless Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 St...

Страница 4: ...ution pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 05 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buff...

Страница 5: ...g against interferences from electrical noise or humidity The flat glass tip allows for direct pH measurement and prevents mash and cooled wort solids from collecting on the surface The extractable cl...

Страница 6: ...d battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Nav...

Страница 7: ...en the tester is in discoverable mode it will appear in the list of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the...

Страница 8: ...e in the second buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 After the second point has been stored CAL SAVE...

Страница 9: ...should be kept hydrated When not in use add a few drops of storage solution to the storage cap If storage solution is not available pH 4 01 or pH 7 01 buffer can be used Note Do not store the electrod...

Страница 10: ...RGA number from the Technical Service department and then send it with shipping costs prepaid When shipping any product make sure it is shipped back in compliance with shipping regulations thoroughly...

Страница 11: ...vide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not insta...

Страница 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810312 Printed in ROMANIA...