Chapter 14
FPD Troubleshooting
Maintaining a Flame Photometric Detector
Maintenance and Troubleshooting Manual
311
The following values of flow rate are recommended:
Table 14-4.
Gas
Capillary columns
Packed columns
Carrier
1÷3 mL/min
30÷50 mL/min
Hydrogen
85÷100 mL/min
100÷120 mL/min
Air
100÷120 mL/min
110÷135 mL/min
Recommended gas flow rates for FPD
The optimum air flow rate should be found experimentally, setting a proper value
for the hydrogen and performing some analyses on a standard mixture at different
air flow rates.
WARNING! It is the customer’s responsibility to ensure compliance with all local safety
regulations concerning gas supplies.
Hydrogen is a potentially dangerous gas. Mixed with air it may give rise to an
explosive mixture. The use of hydrogen requires the operator’s extreme caution due
to the risks involved. For further details concerning hydrogen, please refer to