Getting Started
R&S
®
ZNA
78
User Manual 1178.6462.02 ─ 20
errors, the accuracy requirements of the measurement, on the test setup and on the
types of calibration standards available.
The following example requires a calibration kit with a male Through standard with
known transmission characteristics for the related test port connector type and gender.
With a single Through, it is possible to perform a transmission normalization, compen-
sating for a frequency-dependent attenuation and phase shift in the signal paths.
Due to the R&S
ZNA's calibration wizard, calibration is a straightforward, guided proc-
ess.
1. Replace the DUT by the Through standard of your calibration kit. Make sure to dis-
connect all calibration units.
2. Select Channel – [Cal] > "Configure/Start Calibration..." to open the calibration
setup dialog.
3. "Delete" all except the first row in the "Cal Settings" table. If the table is empty, use
"Add" to create a calibration setup.
4. Tap on the "Calibration" cell of the first row in the "Cal Settings" table to open the
"Define Calibration" dialog.
5. Select port 1 (P1) and port 2 (P2) and the calibration type "Trans Norm".
6. Select "OK" to apply your settings and return to the calibration setup dialog "Cali-
bration Setting" wizard.
Performing measurements