Command reference
R&S
®
ZNA
1503
User Manual 1178.6462.02 ─ 20
●
The set command tries to join the two calibrations.
If successful, it stores the resulting calibration as <OutputCal> in <LocationOutput-
Cal>.
Parameters:
<FirstCal>
String parameter with the channel name (for <LocationFirst-
Cal>=CHANnel) or the calibration file name (for <LocationFirst-
Cal>=POOL) of the first calibration.
<LocationFirstCal>
CHANnel | POOL
<SecondCal>
String parameter with the channel name (for <LocationSecond-
Cal>=CHANnel) or the calibration file name (for <LocationSe-
condCal>=POOL) of the first calibration.
<LocationSecondCal>CHANnel | POOL
Setting parameters:
<OutputCal>
String parameter with the channel name (for <LocationOutput-
Cal>=CHANnel) or the calibration file name (for <LocationOut-
putCal>=POOL) of the resulting calibration.
Note:
If (<OutputCal>,<LocationOutputCal>) matches one of the
input calibrations, this input calibration is replaced.
<LocationOutputCal> CHANnel | POOL
Return values:
<Joinable>
Manual operation:
See
"Input Calibration1/Input Calibration2/Cal Name"
SYSTem:CORRection:WIZard[:IMMediate]
<Dialogs>
Keysight-compatible command to open the
Tip:
When working with the
Chapter 6.1.3, "GPIB Explorer"
raw mode ("Options">"Raw mode" in the IECWIN32 GUI) before executing this com-
mand.
Setting parameters:
<Dialogs>
MAIN | CKIT
MAIN: open the "Calibration > Start Cal" softtool tab
CKIT: open the "Calibration Kits" dialog
Usage:
Setting only
SYSTem:DATE
<Year>, <Month>, <Day>
The command queries or defines the instrument's current date setting.
The setting command requires administrator rights; refer to
SCPI command reference