Command reference
R&S
®
ZNA
1463
User Manual 1178.6462.02 ─ 20
Parameters:
<SettlingDelay>
Settling delay per sweep point.
Factory default is 10 µs (=
10
-5
s).
Default unit: s
Manual operation:
See
"External Generators > Minimum Settling Delay"
SOURce:POWer:REDuce:SDELay
<SettlingDelay>
Deprecated; superseded by
If power reduction at sweep end is enabled (
ON) this defines the settling delay at subsequent sweep starts, i.e.
the time between power-up and sweep start when a new sweep is requested.
This is a global setting.
Parameters:
<SettlingDelay>
Settling delay
Default unit: s
Example:
See
SOURce:POWer:SWEepend:MODE
<SweepEndModes>
Selects the power mode at sweep end.
This is a global setting.
Parameters:
<SweepEndModes> AUTO | REDuce | KEEP
AUTO: at sweep end restore the power at sweep start
REDuce: at sweep end reduce the output power as if the chan-
nel base power was set to its minimum possible value
KEEP: at sweep end keep the power at its current level
Example:
SOURce:POWer:SWEepend:MODE REDuce
Reduce power at sweep end.
SOURce:POWer:SWEepend:SDELay 0.01
Use a settling delay of 10 ms
SOURce:POWer:SWEepend:MODE AUTO
At sweep end, restore the power at sweep start
Manual operation:
See
SOURce:POWer:SWEepend:SDELay
<SettlingDelay>
If sweep end mode REDuce or KEEP is active (see
) this defines the settling delay at subsequent sweep starts, i.e. the time between
power reset and sweep start when a new sweep is requested.
SCPI command reference