GUI reference
R&S
®
ZNA
700
User Manual 1178.6462.02 ─ 20
Method
Selects the Delta-L+ characterization method to be used. Default is "Method 2L". For
background information, see
Delta-L 4.0 PCB characterization
"Method 1L" handles single-ended and balanced measurement results differently; acti-
vate "1L Diff. Mode" for balanced measurements. For "Method 2L" and "Method 3L"
measurements can also be single-ended or balanced, but the tool does not distinguish
between them.
Remote command:
CALCulate:FMODel:DELT<Ph_pt>:METHod
CALCulate:FMODel:DELT:M1L:DIFFmode
Measurements
Defines whether single-ended or balanced ports are measured for Delta-L characteri-
zations.
Remote command:
CALCulate:FMODel:DELT:MEASurement
Port Order
Declares how the DUT is connected to the measurement system. Choosing the wrong
port ordering for your setup, can result in erroneous results.
"Sequential"
Ports 1, 2 and 3, 4 form input/output pairs
"Odd / Even"
Ports 1, 3 and 2, 4 form input/output pairs
Remote command:
CALCulate:FMODel:DELT<Ph_pt>:PORDer
Frequencies
Delta-L computes the uncertainty in the loss-per-inch calculation at specified frequen-
cies.
The "Frequencies" table is pre-populated with the frequencies recommended by Intel.
If you want to enter your own values, uncheck "Use Default Frequencies" and select
"Edit User Frequencies..." to bring up the "User Frequencies" dialog.
Remote command:
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:USEDefault
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:CURRent?
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:CURRent:COUNt?
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:DEFault?
Offset Embed softtool