Command reference
R&S
®
ZNA
1385
User Manual 1178.6462.02 ─ 20
<Seg>
Sweep segment number. Segment numbers must be sequential.
The specified segment number must be smaller or equal to the
number of existing segments plus 1.
If one or more sweep segments with segment numbers
<Seg>
or larger exist in the current channel, then all these existing seg-
ment numbers are incremented by 1 and the new segment is
inserted as segment no.
<Seg>
.
Note that
<Seg>
defaults to 1, so
[SENSe<Ch>:]SEGMent:INSert
inserts a new segment 1,
shifting the existing segments one position up.
Parameters:
<StartFreq>
Start frequency of the segment. See
Default unit: Hz
<StopFreq>
Stop frequency of the segment; see
Default unit: Hz
<Points>
Number of sweep points in the segment. See
<Power>
Internal source power in the segment. See
Default unit: dBm
<SegmentTime>
Duration of the sweep in the segment. See
DWELl
, this parameter is
replaced by
<MeasDelay>
.
Range:
Depending on other channel settings. AUTO acti-
vates automatic sweep time setting in the segment,
which is equivalent to the minimum sweep time
possible.
Default unit: s
<MeasDelay>
Delay for each partial measurement in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect
SWTime
,
this parameter is replaced by
<SegmentTime>
.
<Unused>
Ignored parameter (for compatibility with R&S ZVR analyzers).
Should be set to the default value 0.
<MeasBandwidth>
IF bandwidth in the segment. See
]SEGMent<Seg>:BWIDth[:RESolution]
Default unit: Hz
SCPI command reference