GUI reference
R&S
®
ZNA
359
User Manual 1178.6462.02 ─ 20
5.2.12.3
Power Added Efficiency (PAE) dialog
The "Power Added Efficiency (PAE)" dialog configures the measurement of the PAE of
an active 2-port device.
PAE plugin
●
The PAE measurement is preinstalled on instruments shipped with FW V2.15 or
higher. For older instruments, it is available as a separate installer that can be
downloaded from the
, and that is supported with FW
V2.15 or higher.
●
The PAE measurement requires an external power supply R&S
HMP2030 that is
connected to the R&S
ZNA via USB.
●
There are no dedicated remote commands for the PAE measurements. Instead,
the measurement parameters are provided as:
–
trace definition parameter "Src Port" in the
–
"Additional Input" string in the
and
Access:
Trace – [Meas] > "S-Params" > "External DLL" > "Power Added Efficiency"
Background information
Chapter 4.7.42, "Power added efficiency"
Meas softtool