Command reference
R&S
®
ZNA
1403
User Manual 1178.6462.02 ─ 20
Example:
*RST
Reset the instrument, activating a frequency sweep with 201
sweep points.
SWE:TIME?
Query total sweep time.
SWE:POIN 2010
Multiply the (default) number of points by 10.
SWE:TIME?
Query total sweep time again. The analyzer estimates a sweep
time that is also multiplied by 10.
Manual operation:
See
[SENSe<Ch>:]SWEep:REVerse
<Boolean>
Reverses the direction of the sweep in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
*RST:
OFF
Manual operation:
See
[SENSe<Ch>:]SWEep:STEP
<StepSize>
Sets the distance between two consecutive sweep points in a (non-segmented) linear
frequency sweep.
Suffix:
<Ch>
.
Channel number
Parameters:
<StepSize>
Frequency step size.
The step size is equal to the current sweep spandivided by the
number of sweep points minus one (see
,
respectively). See also the description of manual control and the
program example below.
Range:
Depends on the current sweep span and the maxi-
mum number of sweep points: "Span Frequency" /
(max. "Number of Points" – 1) ≤ "Freq Step Size" ≤
"Span Frequency"
*RST:
Depends on the analyzer model. The default step
size is equal to the default sweep span of the ana-
lyzer divided by the default number of sweep points
minus one.
Default unit: Hz
SCPI command reference