GUI reference
R&S
®
ZNA
408
User Manual 1178.6462.02 ─ 20
you can use the time gate settings to eliminate unwanted responses in your signal.
After switching back to the frequency domain, you will receive the frequency response
of your DUT without the contribution of the unwanted responses. The time gate is inde-
pendent of the frequency window used to filter the trace before transforming it to time
domain.
Time domain analysis requires option R&S
ZNA-K2. If this option is not installed, the
and "Time Gate" tabs are hidden.
Background information
Refer to
Time Gate
Enables or disables the time gate for the time domain and frequency domain traces.
"Gat" is displayed in the trace list while the time gate is active.
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:STATe
Axis Pair
"Start Stop" lets you define the time gate via its "Start" and "Stop", "Center Span" lets
you define it via its "Center" and "Span" value (in time). The analyzer generates a
warning if the (resulting) time span exceeds the unambiguous range which is given by
Δt = 1/Δf, where Δf is the "Freq Step Size". Simply reduce the time span until the warn-
ing disappears.
Traces softtool