GUI reference
R&S
®
ZNA
702
User Manual 1178.6462.02 ─ 20
Remote command:
CALCulate:FMODel:DELT<Ph_pt>:TCONfig
CALCulate:FMODel:DELT:TDR
Output Settings
Allows you to specify the directory where the Delta-L tool stores its data.
Default is
C:\Users\Public\Documents\Rohde-Schwarz\ZNA\DeltaL
.
If "Timestamp Filenames" is checked, the names of subsequently generated "Test Cou-
pon" files are prefixed with the current date and time.
Remote command:
CALCulate:FMODel:DELT<Ph_pt>:DIRectory
CALCulate:FMODel:DELT<Ph_pt>:DIRectory:DEFaultCALCulate:FMODel:
DIRectory:DEFault
OK
"OK" applies the Delta-L settings and starts the
5.14.11.2
Delta-L measurement
When the Delta-L measurement starts, the analyzer firmware prepares the Delta-L
measurement channel, populates the diagram area with the required measurement
traces, and opens the
Delta-L Measurement dock widget
.
Delta-L Measurement dock widget
The "Delta-L Measurement" dock widget is where coupon measurements are per-
formed and/or corresponding touchstone files are selected.
setting in the "Delta-L Settings" dialog determines whether single-
ended or balanced can be measured. "No ports available!" in the dock widget and "No
Trace" in the diagram area indicates that no ports of the selected type are configured.
Access:
●
Channel – [Offset
Embed] > "Delta-L" > "Delta-L Measurement"
●
Offset Embed softtool