Command reference
R&S
®
ZNA
1144
User Manual 1178.6462.02 ─ 20
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:EFEed
<TraceName>
CALCulate<Ch>:PARameter:SDEFine
<TraceName>
)
to a diagram area
<Wnd>
to assign the trace to a diagram area using a numeric suffix
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:OFFSet
command).
Tip:
You can open the "Trace Manager" dialog to obtain an overview of all channels
and traces, including the traces that are not displayed.
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
).
<WndTr>
Trace number. This suffix is ignored; the trace is referenced by
its name.
Setting parameters:
<TraceName>
String parameter for the trace name, e.g. 'Trc4'.
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC:EFE 'CH4TR1'
Display the generated trace in diagram area no. 2. No trace
number is assigned.
Usage:
Setting only
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:FEED
<TraceName>
CALCulate<Ch>:PARameter:SDEFine
) to a diagram
area, using the
<WndTr>
suffix, and displays the trace. Use
WINDow<Wnd>]:TRACe<WndTr>:EFEed
to assign the trace to a diagram area with-
out using a numeric suffix.
Tip:
A trace can be assigned to a diagram only once. If a attempt is made to assign the
same trace a second time (e.g. by typing
DISP:WIND2:TRAC8:FEED 'CH4TR1'
after executing the program example below) an error message -114,"Header suffix out
of range" is generated. You can open the "Trace Manager" dialog to obtain an overview
of all channels and traces, including the traces that are not displayed.
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
).
<WndTr>
Trace number used to distinguish the traces of the same dia-
gram area
<Wnd>
.
Parameters:
<TraceName>
String parameter for the trace name, e.g. 'Trc4'.
SCPI command reference