Command reference
R&S
®
ZNA
1372
User Manual 1178.6462.02 ─ 20
This setting only takes effect, if:
●
Internal Pulse Control is enabled
●
Fixed Duty Cycle is disabled (
[SENSe<Ch>:]PULSe:FXDCycle[:STATe]
OFF)
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Parameters:
<pulse delay on pulse
modulator>
Pulse delay [s]
Default unit: s
Options:
R&S
ZNA<Freq>-B4<Pt> or R&S
ZNA-B91
Manual operation:
See
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:INPut:INVerted[:STATe]
<Boolean>
If External Pulse Control is used, this defines the polarity of the incoming generator
pulses at PuMo In/Out <Pt>.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Parameters:
<Boolean>
*RST:
OFF (0)
Options:
R&S
ZNA-B91
Manual operation:
See
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:OUTPut:INVerted[:STATe]
<Boolean>
If pulse generator <Pt> is output to PuMo In/Out <Pt> (see
PORT<Pt>:EXTernal:OUTPut[:STATe]
), this defines the polarity of the outgoing
pulses.
Suffix:
<Ch>
.
Channel number
<Pt>
Port number
Parameters:
<Boolean>
*RST:
OFF (0)
Options:
R&S
ZNA-B91
Manual operation:
See
SCPI command reference