Limit check and display of limit lines are independent of each other:
–
The limit lines can be displayed, no matter if the limit check is enabled.
–
If "Limit Check" is enabled, the limits are checked, no matter if the limit lines
are displayed.
–
The limit check can even be enabled, if no limit lines are defined. In this case,
the info field displays "No limit defined!" and the limit check always passes.
●
Limit lines are defined for a particular trace format. However, the limit check is per-
formed irrespective of the current trace format. The info field indicates the correct
"PASS"/"FAIL" state and limit violations are visualized on the trace (if any).
Figure 5-15: Limit line violations in complex trace formats
Remote command:
CALCulate<Chn>:LIMit:STATe
CALCulate<Chn>:LIMit:LOWer:STATe
CALCulate<Chn>:LIMit:UPPer:STATe
CALCulate<Chn>:LIMit:FAIL?
CALCulate:LIMit:FAIL:ALL?
CALCulate<Chn>:LIMit:STATe:AREA
Limit Fail Beep
Activates or deactivates the fail beep. The fail beep is a low-tone acoustic signal that is
generated each time the analyzer detects an exceeded limit. No fail beep can be gen-
erated if the limit check is switched off.
Note:
The R&S
ZNA does not have a built-in audio device and loudspeaker. To hear
these sounds, connect a USB audio device to the instrument or operate it via remote
desktop.
Remote command:
CALCulate<Chn>:LIMit:SOUNd[:STATe]
Clear Test
Resets the limit check results.
Remote command: