[SENSe<Ch>:]CORRection:COLLect:DETector
<CalName>[, <CalDetectorType>]
Selects the detector type during manual calibration <CalName>.
Suffix:
<Ch>
.
Channel number
Parameters:
<CalName>
Name of a manual calibration previously created e.g. using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
<CalDetectorType>
NORMal | AVERage
AVERage
Average detector, recommended for channels measuring traces
using detectors other than the NORMal one
Automatically selected for noise figure calibrations (see
[SENSe<Ch>:]CORRection:COLLect:NFIGure:STARt
NORMal
Normal (sample) detector, recommended for all other applica-
tions
*RST:
NORMal
Manual operation:
See
[SENSe:]CORRection:COLLect:FIXTure:LMParameter[:STATe]
<Boolean>
Selects an Auto Length (and Loss) calculation or a Direct Compensation.
Parameters:
<Boolean>
1 - Auto Length (and Loss), depending on the last
[SENSe<Ch>:]CORRection:COLLect:FIXTure:
LMParameter:LOSS[:STATe]
setting
0 - Direct Compensation
*RST:
1
Example:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:
Manual operation:
See
[SENSe:]CORRection:COLLect:FIXTure:LMParameter:LOSS[:STATe]
<Boolean>
Selects an Auto Length or an Auto Length and Loss calculation.
Parameters:
<Boolean>
1 - Auto Length and Loss
0 - Auto Length, no loss
*RST:
1
Example:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:
SCPI command reference