Command reference
R&S
®
ZNA
1236
User Manual 1178.6462.02 ─ 20
Parameters for setting and query:
<Type>
Name of the converter type, whose source multiplier is set/quer-
ied.
Options:
R&S
ZNA-K8
Manual operation:
See
[SENSe:]CONVerter:DEFinition:SPOWer
<Type>, <SourceNomPower>
[SENSe:]CONVerter:DEFinition:SPOWer?
<Type>
Sets/queries the preferred input power at the RF In port of the related converter type.
Parameters:
<SourceNomPower> Default unit: dBm
Parameters for setting and query:
<Type>
Name of the converter type, whose preferred RF input power is
set/queried.
Options:
R&S
ZNA-K8
Manual operation:
See
[SENSe:]CONVerter:DEFinition:SPOWer:OFFSet
<Type>,
<SourceNomPowerCalOffset>
[SENSe:]CONVerter:DEFinition:SPOWer:OFFSet?
<Type>
Sets/queries the cal power offset of the source port that is connected to the RF In port,
when converters of the related type are configured.
See
SOURce<Ch>:POWer<PhyPt>:CORRection:LEVel:OFFSet
and
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>:LEVel:OFFSet
Parameters:
<SourceNomPowerCalOffset>
Default unit: dBm
Parameters for setting and query:
<Type>
Name of the converter type, whose cal power offset is set/quer-
ied.
Options:
R&S
ZNA-K8
[SENSe:]CONVerter:SPLitter<Port>:LOPort
<Local>[, <LocalIndex>]
Defines the source of the LO signal distributed via splitter <Port>.
"Splitter 2" is only useful, if 2 independent 2-port converter measurements shall be per-
formed.
Suffix:
<Port>
.
Splitter number
Parameters:
<Local>
NONE | RLO | PORT | GENerator
SCPI command reference