GUI reference
R&S
®
ZNA
495
User Manual 1178.6462.02 ─ 20
Any generator power calibration or attenuation of the active source step attenuators is
included.
Remote command:
CW Mode
sweeps, are performed at constant frequency and stim-
ulus power, which can be specified in the "Stimulus" settings (see
The measurement is triggered according to the current trigger settings (see
on page 504). Each trigger event triggers the first partial mea-
surement of a measurement point. The time interval between two consecutive mea-
surements depends on the trigger settings and the sweep parameters (especially the
number of points). Any trigger mode is allowed.
A "CW Mode" sweep corresponds to the analysis of a signal over the time with a time
scale and resolution that is determined by the trigger events. In a Cartesian diagram,
the measurement result is displayed as a trace over a linear time scale (like, e.g., in an
oscilloscope). The diagram is similar to the "Time" diagram. The following example
shows a "CW Mode" sweep with a DUT that does not markedly change its transmis-
sion characteristics over the time.
Tip: Sweep time
Sweep Softtool