Command reference
R&S
®
ZNA
936
User Manual 1178.6462.02 ─ 20
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Method>
M1L | M2L | M3L
One, two, or three line/length analysis
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:DELT<Ph_pt>:PORDer
<PortOrder>
Declares how the DUT is connected to the measurement system. Choosing the wrong
port ordering for your setup, can result in erroneous results.
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<PortOrder>
ODDeven | SEQuential
ODDeven
Ports 1, 3 and 2, 4 form input/output pairs
SEQuential
Ports 1, 2 and 3, 4 form input/output pairs
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:DELT<Ph_pt>:RUN
Runs the AITT-DL tool with the collected measurement data and the configured Delta-L
settings.
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage:
Event
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:DELT<Ph_pt>:SWEep:CONTrol
<SweepControl>
Tells the firmware to use either default values for frequency step size and IF band-
width, or the settings of the active channel for the Delta-L measurement.
The resulting frequency step size and IF bandwidth can be queried using
FMODel:DELT<Ph_pt>:SWEep:CONTrol:STEP?
FMODel:DELT<Ph_pt>:SWEep:CONTrol:IFBW?
SCPI command reference