GUI reference
R&S
®
ZNA
443
User Manual 1178.6462.02 ─ 20
The "Define Ripple Test" dialog contains a table to edit the individual ranges of the rip-
ple check ranges. The buttons below the table extend, shorten, or reorder the range list
and save/recall ripple test data.
Range List
Defines the individual ripple limit ranges.
The table contains an automatically assigned current number for each range plus the
following editable columns:
●
"Range On/Off" enables or disables the ripple limit check in each range. Disabling
the ripple limit check does not delete the range but hides the entry in the info field.
●
"Start Stimulus" is the smallest stimulus (x-axis) value of the range.
●
"Stop Stimulus" is the largest stimulus (x-axis) value of the range.
●
"Ripple Limit" is the maximum allowed difference between the largest and the
smallest trace value in the range.
The ripple limit range is displayed as two parallel, horizontal lines in the diagram. "Stop
Stimulus" - "Start Stimulus" is the length of both lines (if the range is within the sweep
range); "Ripple Limit" is their vertical distance. See
Chapter 4.4.1.2, "Rules for ripple
Remote command:
CALCulate<Chn>:RIPPle:SEGMent<Seg>[:STATe]
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:STARt
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:STOP
CALCulate<Chn>:RIPPle:SEGMent<Seg>:LIMit
CALCulate<Chn>:RIPPle:SEGMent:COUNt?
Add / Insert / Delete / Delete All / Align All
The first four buttons below the range list extend, shorten, or reorder the list.
●
"Add" adds a new range to the list. The new range is inserted after the previously
selected range. The current range numbers are adapted; the start and stop stimu-
lus values are set so that an overlap is avoided. Moreover, the ripple limit is estima-
ted according to the measured ripple of the trace in the created range. The ana-
lyzer places no restriction on the number of ranges assigned to each trace.
●
"Insert" adds a new range before the active range (marked by a blue background in
the first column of the range list). The new range extends from the "Stop Stimulus"