Command reference
R&S
®
ZNA
1405
User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]SWEep:TIME:AUTO
<Boolean>
When enabled, the (minimum) sweep time is calculated internally using the other chan-
nel settings and zero delay (
Note:
The automatically calculated sweep duration is ignored for the sweep types
).
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
ON | OFF - turn the automatic calculation of the sweep time on
or off.
OFF is also set if the sweep duration or delay is set explicitly
using
*RST:
ON
Example:
SWE:TIME 1
Set a total sweep time of 1 s.
SWE:TIME:AUTO?
A query returns the value
1
.
Manual operation:
See
[SENSe<Ch>:]SWEep:TIME:MODE
<Algorithm>
Defines how the firmware calculates the "hardware settling wait times" in channel
<Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Algorithm>
AUTO | MONO | CONV | SLOW
AUTO
Let the firmware choose the suitable algorithm, based on the
channel's current measurement task
MONO
Manually (and statically) select the algorithm for mono-frequent
measurements
CONV
Manually (and statically) select the algorithm for frequency-con-
verting measurements
SLOW
Manually (and statically) select the algorithm for other "slow"
measurements (e.g. involving power meters)
*RST:
AUTO
Manual operation:
See
SCPI command reference