Command reference
R&S
®
ZNA
1455
User Manual 1178.6462.02 ─ 20
ON | 1
Only those frequencies are calibrated that are relevant for this
port. This results in shorter calibration times.
*RST:
0
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Setting only
SOURce<Ch>:POWer<PhyPt>:CORRection:NREadings
<NoCalSweeps>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels, ports and external generators.
Equivalent command:
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the limit
applies to all sources.
<PhyPt>
Calibrated port number. This parameter is ignored; the limit
applies to all sources.
Parameters:
<NoCalSweeps>
Number of readings
Range:
1 to 100
*RST:
2
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
"Flatness Cal – Max Iterations"
SOURce<Ch>:POWer<PhyPt>:CORRection:OSOurces[:STATe]
<Boolean>
Switches off all other sources during the calibration sweep for channel
<Ch>
.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number. This parameter is ignored; the setting is
valid for all sources.
Parameters:
<Boolean>
ON - other sources not necessarily switched off.
OFF - other sources switched off during the calibration sweep.
*RST:
ON
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See