Command reference
R&S
®
ZNA
1317
User Manual 1178.6462.02 ─ 20
Parameters:
<Wave>
Identifier for the calibrated wave:
'A<n>' or 'AWAV<n>' denote correction data for the reference
wave a<n>, where <n> corresponds to the port number.
'B<n>' or 'BWAV<n>' denote correction data for the reference
wave b<n>.
<CorrData>
Power correction values either in ASCII or block data format,
depending on the current
Example:
*RST; :SWE:POIN 10
Reset the instrument and reduce the number of sweep points to
10.
CORR:POW:ACQ AWAVe,1,PORT,1
Perform a receiver power calibration of the wave a
1
using port 1
as a source port, assuming that the transmitted source power a
1
is correct. No external cabling is needed.
CORR:POW:DATA? 'AWAV1'
Query the correction values. The analyzer returns 10 comma-
separated real numbers.
CORR:POW:DATA 'AWAV1', 1, 2, 3, 4, 5, -6, -7,
-8, -9, -0
Replace the correction values by ten (new) numbers.
Manual operation:
See
[SENSe<Ch>:]CORRection:POWer<PhyPt>[:STATe]
<Boolean>
Enables or disables the receiver power calibration for channel
<Ch>
and for the
received waves b
<PhyPt>
. The setting command is disabled unless the received waves
have been power calibrated (
[SENSe<Ch>:]CORRection:POWer<PhyPt>:
BWAVe, ...
). The query always returns a result.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the receiver power calibration
for the received waves b
<PhyPt>
.
*RST:
OFF
SCPI command reference