Command reference
R&S
®
ZNA
1299
User Manual 1178.6462.02 ─ 20
<TestPort1>
First calibrated test port number. For an n-port calibration type, n
port numbers must be specified. If more than n numbers are
defined, the spare numbers (the last ones in the list) are ignored.
Entering less than n numbers causes an error message. For a
one path two port calibration (
OPTPort
), the first port no.
denotes the node port (fully calibrated port).
<TestPort2>
Second calibrated port number.
<TestPort3 or
AdapterCalKit>
Either the third calibrated port number, or - for Adapter Removal
calibrations only - the file path (relative to the current directory)
where the analyzer shall store the adapter characterization
obtained during the calibration.
The latter is optional. If a file path is provided the characteriza-
tion file (in Touchstone s2p format) is generated when the cor-
rection terms are finally calculated (
]CORRection:COLLect:SAVE:SELected[:DUMMy]
<TestPorts>
More ports to be calibrated.
Example:
:SENSE1:CORRECTION:COLLECT:METHOD:DEFine
'Test', ARTosm, 1,2, 'Adapter.s2p'
Initiate an Adapter Removal calibration for ports 1 and 2 on
channel 1 and store the adapter characterization in
<MMEMORY:CDIR?>\Adapter.s2p
.
For another example see
Manual operation:
See
The supported calibration types are listed below.
Parameter
Type ("Calibration Presetting" dialog)
ARTosm
Adapter Removal
REFL | RSHort
Refl Norm Open | Refl Norm Short
FOPort
Full One Port ("Refl OSM")
FTRans | RTRans | FRTRans
Trans Norm - Forward | Trans Norm - Reverse | Trans Norm Both
"Forward" ("Reverse") means that the port with the smaller (larger)
port number serves a a drive port.
OPTPort
One Path Two Ports
TOSM
TOSM
UOSM
UOSM
VUOSm
UOSM Vector Mixer
TRL | TOM | TSM | TRM | TNA
TRL | TOM | TSM | TRM | TNA
ARPTosm | PTOSm | PUOSm | PTRL
| PTOM | PTSM | PTRM | PTNA
One of the previous calibration types in combination with a power
calibration (for SMARTerCal).
VP
VPUosm
An alias for PUOSm
SCPI command reference