Basic mode
In basic mode, the validation runs automatically, with autodetected ports and the set-
tings specified in
(default settings after a system [Preset]).
Use "Detect Port Assignment" to detect the current port assignment (after reconnecting
the calibration unit).
Manual port assignment is not supported.
Validate
Runs the validation for the current port assignment. The validation proceeds through
the connected ports p
∈
{1, ..., 4} in ascending order, measuring S
pp
for the Open,
Short and Match standard.
●
"PASS": all measured reflection parameters are within the specified
, compared to the selected reference
●
"FAIL" otherwise.
[In this case, the validation stops at the first failed reflection measurement.]
Remote command:
Advanced Mode
Switches to
.
Advanced mode
In advanced mode you can configure the validation settings (
) and run or rerun particular measurements (
,
Cal softtool