GUI reference
R&S
®
ZNA
547
User Manual 1178.6462.02 ─ 20
left: = source and meas. port identical
right = source and meas. port different
Port Overview
The dialog shows all receiver ports of the network analyzer. The selected port is dis-
played with the current cal power settings (see
Chapter 5.11.3.2, "Cal Power Config
on page 574); moreover, a circuit diagram visualizes the purpose of the mea-
surement receiver calibration.
"Source Port" defines the type of measurement receiver calibration:
●
If the source port is equal to the calibrated port, the measurement receiver is cali-
brated by the wave that is reflected back by a connected Open or Short standard.
Connect the Open or Short standard to the calibrated port; no additional external
test setup is required.
●
If the source port and the calibrated port are different, the measurement receiver is
calibrated by the wave generated at the source port. Connect the source port to the
calibrated port, including any external devices that you used for the source power
calibration.
For the source port, a source power calibration must be active.
Remote command:
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
Cal softtool